COULOSCOPE® CMS2 and CMS2 STEP
Our coulometry specialist.
Benchtop instrument for measuring coating thicknesses and electrochemical potentials using the coulometric method.
Key Benefits
- Optimal measurement concept.: Predefined measurement tasks for different coating systems
- Easy calibration.: Achieve the highest level of accuracy
- Individually expandable.: Extensive accessories enable practical work and safe storage
- Intuitive operation.: Color display and graphically supported user guidance
Description
Benchtop instrument for measuring coating thicknesses and electrochemical potentials using the coulometric method.
Above a certain coating thickness, using X-ray fluorescence to measure coating thicknesses non-destructively will reach its limits. This is where the COULOSCOPE® CMS2 comes into play. Use the coulometric method to measure the coating thickness of almost all metal coatings precisely and reliably by electrolytic layer detachment. This easy-to-use benchtop device offers maximum flexibility, as it can be used for a wide range of coating-substrate combinations.
Predefined measurement tasks for different coating systems
Achieve the highest level of accuracy
Extensive accessories enable practical work and safe storage
Color display and graphically supported user guidance
Exact coating thickness measurement of almost any coating-substrate-combination
Graphic representation of the voltage curve at the measuring cell
Almost 100 predefined measuring tasks for different coating systems (such as zinc on iron or nickel on brass)
Deplating the coating according to DIN EN ISO 2177
Simple data transfer via USB interface
Measuring range: Depending on the coating-base material combination and the removal speed 0.02 - 50 µm
Measured value memory: 3,000 in 50 applications
Functions for STEP test measurement in accordance with ASTM B764 - 04 and DIN EN 16866
Simple selection of the detachment speed (0.1 - 50 µm/min) and the detachment area (0.6 - 3.2 mm Ø)
Measuring method: Coulometry by anodic stripping
Do you have other applications? Then contact us!
Features
- Graphic representation of the voltage curve at the measuring cell
- Almost 100 predefined measuring tasks for different coating systems (such as zinc on iron or nickel on brass)
- Deplating the coating according to DIN EN ISO 2177
- Simple data transfer via USB interface
- Measuring range: Depending on the coating-base material combination and the removal speed 0.02 - 50 µm
- Measured value memory: 3,000 in 50 applications
- Functions for STEP test measurement in accordance with ASTM B764 - 04 and DIN EN 16866
- Simple selection of the detachment speed (0.1 - 50 µm/min) and the detachment area (0.6 - 3.2 mm Ø)
- Measuring method: Coulometry by anodic stripping
- Predefined measuring tasks for different layer systems
- Measuring of the entire layer structure
- Simple calibration for a maximum degree of accuracy
- Measuring the coating thickness of almost all metallic single and multilayer coatings on almost any base material
- Quality control of electroplated coatings
- Monitoring the residual tin layer thickness on PCBs
- STEP test measurement: measuring single layer thicknesses and potential differences in multiple nickel layers on copper, iron, aluminum or plastic substrates (ABS)